The aim of a first session is to learn what your unit does, not to find its limits. Use two self-adhesive pads on one thigh. It's a big muscle, well away from the chest, and easy to see and reach.
Read E-Stim Safety first. If any of the "who shouldn't use it" points apply to you, stop there.
Before you start#
- Read the unit's own manual. Where it disagrees with anything here, follow the manual.
- Fit fresh batteries, or make sure it's charged and unplugged from the charger.
- Check the leads for kinks, cracks or bare wire.
- Pick a time when you're awake, sober and won't be interrupted.
- Make sure the skin is clean, dry and unbroken. Lotion and oil stop pads sticking and block the current.
Setting up#
- Unit off, levels at zero.
- Plug the lead into channel A and push the two pins fully into the two pads.
- Stick both pads on the front of one thigh, a hand's width apart. Press them down all over, edges included.
- Check the pads aren't touching each other.
- Choose the plainest mode the unit has, usually a steady or "continuous" one.
Bringing it up#
- Switch on. You should feel nothing yet.
- Raise the level one step at a time, pausing for a few seconds between steps.
- Stop at the first faint tingle. That's your threshold.
- Go a little higher until the feeling is clear but comfortable, then leave it there for a minute.
It's normal for the sensation to fade after a while as the nerves get used to it. Resist the urge to keep chasing it upwards. Changing the pattern works better than adding power.
Things to try#
Turn the level down a few steps before each change, then bring it back up.
- Change the pulse rate. Slow feels like separate taps. Fast blurs into a smooth buzz.
- Change the pulse length, if your unit lets you. Longer pulses feel deeper and stronger at the same level.
- Try another mode. Waves and ramps feel very different from steady pulses.
- Move the pads. Turn the level to zero first. Closer together feels shallower, further apart feels deeper.
What's normal and what isn't#
Normal: tingling, tapping, buzzing, a pulling or squeezing feeling, and the muscle twitching at higher levels.
Stop and check: a sharp sting or a hot, prickly spot under part of a pad. That usually means poor contact. Turn down, switch off and re-seat the pad. See E-Stim Troubleshooting.
Stop for the day: pain, cramp, feeling dizzy or unwell, or skin that looks marked or blistered.
Finishing#
- Turn the level down to zero.
- Switch the unit off.
- Peel the pads off by the pad, not the wire, and put them back on their backing sheet.
- Look at the skin. A little pinkness that fades in a few minutes is common. A mark that stays or feels sore is a mild burn: stop using that spot until it's healed and work out what caused it before you try again.
Keep the first session short. Ten to fifteen minutes is plenty.
Next steps#
- Pulses, Frequency and Waveforms explains what you just felt.
- Where Electrodes Can and Can't Go before you try anywhere else.
- Electrode Types when you want to move beyond pads.